Easy to use, compact design water bath
Features BM100/110/200/210
- Analog set up system
- Thermometer is included to verify actual temperature
- Protected water tank prevents burns caused by contact
- Equipped with a drain (BM200/210)
BM302A/312A
- White LED digital display, key entry
- Flat-shaped bath with no heater or sensor inside makes for ease in cleaning
- Full range of safety functions
BM401/410
- Digital temperature setting by ▲▼ keys
- Protected water tank prevents burns caused by contact
- Equipped with a drain
BM500/510
- Digital temperature setting by ▲▼ keys
- Removable water tank for convenient cleaning and changing of water
- Heater situated outside the water tank
- Exclusive connection for bath operational setting function with RE601/801 rotary evaporator
Models | BM100 BM110 |
BM200 BM210 |
BM302A BM312A |
BM401 BM410 |
BM500 BM510 |
|
Operating Temp. Range *¹ | Room temp. +5~95°C *¹ | Room temp. +5~95°C *¹ | Room temp. +10~90°C | Room temp. +5~95°C *¹ | Room temp. +5~90°C | |
Temp. Adjustment Accuracy*² | ±2°C (at 60°C) | ±2°C (at 60°C) | ±1.0°C | ±1°C (at 60°C) | ±1.5°C (at agitation) | |
Temp. Control System | ON/OFF control | ON/OFF control | PlD control | PlD control by microprocessor | PlD control by microprocessor | |
Temp. Setting / Display Method | Analog setting (Glass thermometer indication) | Analog setting (Glass thermometer indication) | White LED digital display, key entry, minimum digit of 1°C | Digital setting by ▲▼ keys | Digital 7 segment LED Digital setting by ▲▼ keys | |
Operation Function | Operation at fixed point | Operation at fixed point | Operation at fixed point | Operation at fixed point | Fixed temperature, quick auto stop, auto stop, auto start | |
Additional Functions | – | – | – | Temp. preset function (preset / select one temp.) | Keypad lock, maintenance function (RE signal transmission and reception), calibration off-set, power failure recovery | |
Heater | SUS316 pipe heater 500W | SUS316 pipe heater 1kW | 1000W aluminum sheathed heater | SUS316 pipe heater 1kW | 1kW (100V) / 1.44kW (120V) 1kW (200V) / 1.44kW (220V) |
|
Sensor | Liquid expansion type | Liquid expansion type | Pt100Ω | Pt100Ω | K thermocouple | |
Safety Device | Bath protection cover | Bath protection cover | Bath protection cover | Bath protection cover | Bath protection cover (ABS heat-resistant resin) | |
Overcurrent protection (fuse: 7A), thermal fuse | Overcurrent protection (fuse: 15A), thermal fuse | Automatic overheat prevention, independent overheat prevention (fixed temp.), temperature upper limit error, Overcurrent protection fuse, Overshoot alert, Auto resume (selectable), calibration offset, 2A service outlet (for BM302A) |
Self-diagnostic functions (automatic overheat preven- tion, sensor trouble, triac short circuit, heater disconnection, main relay failure), circuit protector, thermal fuse | Self-diagnostic functions (automatic overheat prevention, sensor trouble, triac short circuit, heater disconnecton, main relay failure), circuit protector, thermal fuse, micro switch to detect heating without water | ||
Water Tank Capacity | ~4L | ~7L | ~5L | ~7L | ~4L | |
Water Tank Dimensions | ø200 x D120 mm | ø250 x D150 mm | ø262 x D286 | ø250 x D150 mm | Max. ø240 x bottom ø165 x D122 mm | |
External Dimensions WxDxH | 240 x 300 x 150 mm | 310 x 360 x 230 mm | 240 mm (H) | 310 x 360 x 230 mm | 340 x 349 x 231 mm | |
Weight | ~3.5kg | ~6kg | ~4.5kg | ~7kg | ~5.5kg | |
Power Source (50/60 Hz) | AC115V 4.5A paddle switch AC220V 2.3A paddle switch |
AC115V 9A AC220V 4.5A |
AC100-115V, 10-12A (service outlet excluded) AC200-230V, 5-6A |
AC115V 9A AC220V 4.5A |
AC100V~120V 12.5~10.5A AC200~240V 6.5~5.5A |