Control range
Performance
- Microprocessor PID control / Temperature calibration / Automatic tuning.
- Advanced performance of low and high temperature and humidity control, achieved by adapting the humidification tank exterior to the chamber, added benefit for long term test.
- Water recycling system reduce water consumption for longer operation.
- Air-tight door closure with perfect sealed packing.
- Heat-resistant silicone packing completely blocks high heat leakage from the inner chamber.
- Two-point door latch also enhances tightness to door closure.
- Energy efficiency is achieved by optimization of the insulation system.
- Even if the unit is operating at max. temperature, the surface of the unit doesn’t allow high temperatures of over 60℃.
Safety
- Built in door lock with a key.
- Automatic shut off after overheat alarm.
- Low and empty water level alarm.
- Power supply leakage breaker.
- Over current protection.
- Open door alarm.
Convenience
- Three different volume sizes are available to suit any users’ needs.
- Corrosion resistance stainless steel chamber and electric polished humidification tank.
- Water supply during operation is also available for long term test purposes.
- Direct water system provides user great convenience when performing long-term tests. (optional
- Clear observation of samples.
- Tempered glass door with built in heating wire.
- Energy efficient, long life span LED lighting.
- Wider view, inner glass door is available for monitoring of samples. (optional)
- Cable port for introduction of external probes or wires.
- 50mm diameter cable port comes standard on the left side of the unit.
- Additional 50mm or 80mm diameter cable ports, custom ordered. (optional)
- Casters for easy mobility during installation or relocation.
- User friendly LCD controller.
- Colorful interactive touch type LCD controller.
- Set point temperature and present temperature are clearly shown together on display.
- Sufficient programmable control support.
- Operating progress can be shown in graph mode for better data understanding.
- Automatic start/stop can be programmed for unattended operations.
- Display indicates abnormal temperature changes or power failures.
- Selective bilingual language support.
(Korean / English or English / Chinese, selectable only when placing an order) - Temperature and other operating parameters can be recorded using a six channel digital recorder.(optional)
- Computer interface.
- RS-485 communication port (optional) enables connection up to maximum 32 units. Simultaneous monitoring and operating of multiple units is an available feature.
- Program patterns are simple to set with the use of our software.
- Easy to check status of real-time operations. Graph data supports easy understanding.
- Storage and documentation of programs.
- Data with saving capabilities in excel and text format.
- Storage of program information, backup of value settings, recovery, and temperature values.
- RS-232 communication port as standard offer.
Description | No | |
Programmable pattern capacity | 100 | |
Pattern repeat time | 999 | |
Part repeat time | 255 | |
Max. segments per a pattern | 100 | |
Available max. segments* | 2000 | |
Programmable process time per a segment | 0 to 99 hr. 59 min. |
* Even though the max. segments per pattern are 100, the total available segments are up to 6000 not 10000, therefore arrangement of the number of segment and patterns will need to be done properly.
Models | TH-DG-150 | TH-DG-300 | TH-DG-400 | |||||
Chamber volume (L / cu ft) | 150 / 5.30 x 2 chambers | 288 / 10.17 x 2 chambers | 384 / 13.56 x 2 chambers | |||||
Controller | PID control (LCD Type) | |||||||
Temperature | Range (°C / °F) | -5 to 100 / 23 to 212 without humidity, 15 to 90 / 59 to 194 with humidity | ||||||
Fluctuation 1) (±°C / °F) at 40℃ / 75%RH | 0.3 / 0.55 | |||||||
Variation 1) (±°C / °F) at 40°C / 75%RH | 0.5 / 0.90 | |||||||
Time | Heating – without humidity (min.) | 85 | ||||||
Cooling – without humidity (min.) | 100 | |||||||
Humidity | Range (%RH) | 15 to 90 ※ See Feature page for more detailed humidity range check. |
||||||
Fluctuation1) (±%RH) at 75%RH / 40°C | 0.8 | |||||||
Variation1)(±%RH) at 75%RH / 40°C | 3.0 | |||||||
Refrigeration | System | Air – cooled | ||||||
Refrigerant | R- 507A | |||||||
Water supply | Capacity | 20 / 0.7 | 24 / 0.84 | |||||
Water quality | pH6.2 to 7.2 / Electrical conductivity 20㎲ /cm to below | |||||||
No. of wire shelves (standard / max.) | 2 / 7 | 2 / 8 | 2 / 8 | |||||
Distance of between shelves (mm / inch) | 60 /2.4 | 60 /2.4 | 60 /2.4 | |||||
Max. load per shelf (kg / lbs) | 25 / 55.1 | 25 / 55.1 | 25 / 55.1 | |||||
Distance of between shelves (mm / inch) | 60 / 2.4 | 60 / 2.4 | 60 / 2.4 | |||||
Permitted total load (kg / lbs) | 75 / 165.3 | 75 / 165.3 | 75 / 165.3 | |||||
Dimension (W×D×H) | Interior (mm / inch) | 600x500x500 / 23.6×19.7×19.7 | 800x600x600 / 31.5×23.6×23.6 | 800x600x800 / 31.5×23.6×31.5 | ||||
Exterior (mm / inch) | 1055x1320x1725 / 41.5×52.0x67.9 | 1255x1470x1925 / 49.4×57.9×75.8 | 1255x1670x1925 / 49.4×65.7×75.8 | |||||
Viewing windows (mm / inch) | 200×330 / 7.9×13 | |||||||
Net weight (kg / lbs) | 420 / 926 | 516 / 1137.6 | 560 / 1234.6 | |||||
Cable port | Basically one hole provided / Ø 50mm (2”), Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”) | |||||||
Interface port | Basically RS-232C, [RS-485 : optional] | |||||||
Electrical requirements | 1ph / 13.6 A x 2ea | 1ph / 17.5 A x 2ea | ||||||
Cat.No. | AAHC6001K | AAHC6011K | AAHC6021K | |||||
Electrical requirements (230V, 60Hz) | 1ph / 13.6 A x 2ea | 1ph / 17.5 A x 2ea | ||||||
Cat.No. | AAHC6002K | AAHC6012K | AAHC6022K |