Performance
General control system
- 5°Cto 50℃(lamp off) / 10°Cto 50°C. (lamp on)
- Max. to 35,000Lux for GC-1000.
Max. to 20,000Lux for GC-300 Models. - 40 to 80% RH for GC-300TLH. (at 20 to 35°C)
50 to 90% RH for GC-1000TLH. (at 20 to 35°C) - Max. 5,000ppm CO2 on/off system. (optional)
- Microprocessor PID control / Temperature calibration / Automatic tuning.
- 10 step programmable temperature, humidity, and illumination profiles and repeatable steps of up to 999 cycles.
Illuminance control system
- Uniformed luminescence distribution.
- Broad distribution of side lamps for hastening the growth of plants.
- High intensity illumination of upper lamp for light efficient and low thermal load. (for GC-1000TLH/1000TL)
- Unique construction for minimization of heat increase from surrounding lamps.
- Tempered glass door blocks heated air from lamps.
- Designed to exhaust heated air through upper vent holes. (for GC-1000TLH)
- Stable and long lasting lighting through introduction of high frequency electronic ballast lamps.
Convenience
- Ergonomically designed door construction.
- Inner glass door with silicone and external door with magnetic sealing for dual airtight packing.
- Well designed providing a smooth open for minimization of damage to plants.
- Wide inner tempered glass door for clear observation of plant growth without affecting inner chamber’s environment.
- Tall plants can be grown inside of the chamber with adjustment of shelve level.
- Maintenance of water supply are easily performed with a front water tank. (GC-1000TLH)
- Water level check indicated by water level bar in the tank.
- Water supply during operation is also available for long term test purposes.
- Adjustable water tank position. (for GC-300TLH)
- Detachable condenser air-filter for easy maintenance of refrigerating efficiency.
- Maintenance of air filter no longer cumbersome with our detachable condenser air filter.
- Casters for easy mobility during installation or relocation.
- Eco-friendly CFC-free refrigerant use.
Safety
- Automatic shut off after overheat alarm.
- Low and empty water level alarm.
- Power supply leakage breaker.
- Over current protection,
- Open door alarm.
Lamps setting
- Control illumination values with the below lamp setting arrangements.
Models | GC-300TL | GC-300TLH | GC-1000TLH | ||||
Chamber volume (L / cu ft) | 300 / 10.6 | 300 / 10.6 | 1000 / 35.3 | ||||
Control system | Microprocessor PID controller | ||||||
Temperature | Range (°C / °F) | 5 to 50 / 41 to 122 – Lamp off | |||||
Fluctuation (±°C / °F) at 25°C | 0.1 / 0.18 – without humidity | ||||||
Variation (±°C / °F) at 25°C | 0.8 / 1.43 – without humidity | ||||||
Illumination | Range (Lux) | 0 to 20,000 | 0 to 20,000 | 0 to 35,000 | |||
Control (steps) | 10 | 10 | 11 | ||||
Lamp | Fl lamp (32W×18ea) | FL lamp (32W×16ea) | |||||
Metal lamp (400W×3ea) | |||||||
Humidity | Range (%RH) | – | 40 to 80 at 20 to 35°C | 50 to 90 at 20 to 35°C | |||
70 to 90 at 36 to 50°C | 60 to 90 at 36 to 50°C | ||||||
Fluctuation (±RH) at 60% RH | – | 3 | – | ||||
Co2(optional) | Range (ppm) | Max. 5,000 | |||||
Sensor | NDIR CO2 sensor | ||||||
Refrigerant (HP) | 1/2 | 1/2 | 3/4 | ||||
Temp. heater power (W) | 750×2ea | 750×2ea | 1,500×2ea | ||||
Humid. heater power (W) | – | 1300 | 1500 | ||||
Dimension (W×D×H) | Interior (mm / inch) | 510×540×1100 / 20×21.3×43.3 | 1200×800×1080 / 47.2×31.5×42.5 | ||||
Exterior (mm / inch) | 700×805×1900 / 27.6×30.1×74.8 | 1410×1070×2150 / 55.5×42.1×84.7 | |||||
Net weight (kg / lbs) | 250 / 551.2 | 550 / 1212.5 | |||||
Electrical requirements (230V) | 60Hz / 12.5A | 50Hz / 12.5A | 60Hz / 16A | 50Hz / 16A | 60Hz / 30A | 50Hz / 30A | |
Electrical requirements (230V) | AAHA1011K | AAHA1012K | AAHA1021K | AAHA1022K | AAHA1031K | AAHA1032K |