Important characteristics
- Temperature range: room temperature plus 7 °C to 60 °C
- Humidity range: up to 95% RH
- Alternative O₂ control ranges: 0.2-20 vol. % O₂ or 10-95 vol. % O₂ (O, GO versions)
- Auto-sterilization with hot air at 180 °C
- Double-pan humidification system with condensation protection
- CO₂ gas-mixing jet with Venturi effect
- Hot-air sterilizable CO₂ sensor with infrared technology
- Display via color LCD monitor
- Lockable door handle
- Tightly-sealed inner door made of tempered safety glass
- Seamless, deep-drawn inner stainless steel chamber
- 2 perforated stainless steel shelves
- Units are stackable with stacking adapter
- Troubleshooting system with visual and audible alarms
- Computer interface: Ethernet
- Internal data recording and USB interface
- Zero-voltage alarm contact
Designation | CB060-230V¹ | CB060-230V-G¹ | CB060-230V-O¹ | CB060-230V-GO¹ | CB060UL-120V¹ | CB060UL-120V-G¹ | CB060UL-120V-O¹ | CB060UL-120V-GO¹ |
Option model | Standard | with 4x divided inner door | with O₂ control | with O₂ control and with 4x divided inner door | Standard | with 4x divided inner door | with O₂ control | with O₂ control and with 4x divided inner door |
Order number | 9040-0088 | 9040-0104 | 9040-0090 | 9040-0106 | 9040-0089 | 9040-0105 | 9040-0091 | 9040-0107 |
Measures | ||||||||
Interior volume [L] | 53 | 53 | 53 | 53 | 53 | 53 | 53 | 53 |
Net weight of the unit (empty) [kg] | 60 | 65 | 63 | 68 | 60 | 65 | 63 | 68 |
permitted load [kg] | 30 | 30 | 30 | 30 | 30 | 30 | 30 | 30 |
Load per rack [kg] | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 |
Doors | ||||||||
Inner doors | 1 | 4 | 1 | 4 | 1 | 4 | 1 | 4 |
Unit doors | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 |
Internal Dimensions | ||||||||
Width [mm] | 400 | 400 | 400 | 400 | 400 | 400 | 400 | 400 |
Height [mm] | 400 | 400 | 400 | 400 | 400 | 400 | 400 | 400 |
Depth [mm] | 330 | 330 | 330 | 330 | 330 | 330 | 330 | 330 |
Outer dimensions | ||||||||
Width net [mm] | 580 | 580 | 580 | 580 | 580 | 580 | 580 | 580 |
Height net [mm] | 720 | 720 | 720 | 720 | 720 | 720 | 720 | 720 |
Depth net [mm] | 545 | 545 | 545 | 545 | 545 | 545 | 545 | 545 |
Wall clearance back [mm] | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
Wall clearance sidewise [mm] | 50 | 50 | 50 | 50 | 50 | 50 | 50 | 50 |
Environment-specific data | ||||||||
Energy consumption at 37 °C [Wh/h] | 80 | 80 | 80 | 80 | 80 | 80 | 80 | 80 |
Fixtures | ||||||||
Number of shelves (std./max.) | 2/3 | 2/2 | 2/3 | 2/2 | 2/3 | 2/2 | 2/3 | 2/2 |
Performance Data Temperature | ||||||||
Temperature range 7 °C above ambient temperature to [°C] | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
Temperature variation at 37 °C [± K] | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 |
Temperature fluctuation at 37 °C [± K] | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 |
Recovery time after 30 seconds door open at 37 °C [min] | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 |
Performance Data Climate | ||||||||
Humidity range [% RH] | 90…95 | 90…95 | 90…95 | 90…95 | 90…95 | 90…95 | 90…95 | 90…95 |
CO₂-Data | ||||||||
CO₂ range [Vol.-% CO₂] | 0…20 | 0…20 | 0…20 | 0…20 | 0…20 | 0…20 | 0…20 | 0…20 |
CO₂ measurement | IR | IR | IR | IR | IR | IR | IR | IR |
CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min] | 5 | 5 | 5 | 5 | 5 | 5 | 5 | 5 |
Electrical data | ||||||||
Rated Voltage [V] | 200…240 | 200…240 | 200…240 | 200…240 | 100…120 | 100…120 | 100…120 | 100…120 |
Power frequency [Hz] | 50/60 | 50/60 | 50/60 | 50/60 | 50/60 | 50/60 | 50/60 | 50/60 |
Nominal power [kW] | 1 | 0 | 1 | 0 | 1 | 0 | 1 | 1 |
Unit fuse [A] | 10 | 10 | 10 | 10 | 16 | 16 | 16 | 16 |
Phase (Nominal voltage) [ph] | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 |
O₂-Data | ||||||||
Standard O₂ control ranges: O₂ range [Vol.-% O₂] | – | – | 0.2…20 | 0.2…20 | – | – | 0.2…20 | 0.2…20 |
O₂ control ranges with option: O₂ range [Vol.-% O₂] | – | – | 10…95 | 10…95 | – | – | 10…95 | 10…95 |
O₂ Recovery time after 30 seconds door open at 5 Vol.-% O₂ [min] | – | – | 8 | 8 | – | – | 8 | 8 |