The most versatile among the cooled incubators for microorganisms: The BINDER KB series cooled incubator controls temperature ranges of -5 °C to 100 °C. With its comprehensive program functions, this cooled incubator offers a wide range of capabilities and delivers reproducible test results.
- Temperature range: -5 °C to 100 °C
- APT.line™ preheating chamber technology
- Up to 30% lower energy consumption compared to the previous model
- Cooling with compressor cooling unit
- Adjustable fan speed
- Controller with time-segment and real-time programming
- Inner door made of tempered safety glass
- 2 stainless steel racks
- Stackable
- Class 3.1 independent temperature safety device (DIN 12880) with visual and acoustic alarm/li>
- Computer interface: Ethernet
Designation | KB115-230V¹ | KB115UL-120V¹ |
Option model | Standard | Standard |
Order number | 9020-0397 | 9020-0398 |
Measures | ||
Interior volume [L] | 115 | 115 |
Net weight of the unit (empty) [kg] | 106 | 106 |
permitted load [kg] | 50 | 50 |
Load per rack [kg] | 20 | 20 |
Doors | ||
Inner doors | 1 | 1 |
Unit doors | 1 | 1 |
Internal Dimensions | ||
Width [mm] | 600 | 600 |
Height [mm] | 480 | 480 |
Depth [mm] | 400 | 400 |
Outer dimensions | ||
Width net [mm] | 835 | 835 |
Height net [mm] | 1025 | 1025 |
Depth net [mm] | 650 | 650 |
Wall clearance sidewise [mm] | 100 | 160 |
Wall clearance back [mm] | 100 | 100 |
Environment-specific data | ||
Energy consumption at 37 °C [Wh/h] | 75 | 75 |
Sound-pressure level [dB(A)] | 49 | 49 |
Fixtures | ||
Number of shelves (std./max.) | 2/5 | 2/5 |
Performance Data Temperature | ||
Temperature range [°C] | -5…100 | -5…100 |
Temperature variation at 37 °C [± K] | 0.2 | 0.2 |
Temperature fluctuation at 37 °C [± K] | 0,1 ± K | 0,1 ± K |
Recovery time after 30 seconds door open at 37 °C [min] | 2 | 2 |
Electrical data | ||
Rated Voltage [V] | 230 | 120 |
Power frequency [Hz] | 50 | 60 |
Nominal power [kW] | 0.7 | 0.7 |
Unit fuse [A] | 10 | 12.5 |
Phase (Nominal voltage) [ph] | 1 | 1 |