- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results.
- Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F).
- MP Controller, 2 programs with 10 sections each or alternatively switch over to 1 program with 20 sections.
- The time interval of single program sections can be adjusted up to a maximum of 99:59 hours or 999:59 hours. This adjustment applies to all program sections.
- Integrated week programm timer with real time function.
- Adjustable time functions and programmable temperature ramp function via program editor.
- Digital temperature setting with an accuracy of one degree.
- Adjustable fan speed.
- Adjustable front ventilation flap slide and rear exhaust ø 50 mm (1.97 inch).
- Elapsed time indicator.
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm.
- RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232 / RS 422 interface converter.
- Units up to 115 liters are stackable.
- 2 chrome-plated racks included.
- BINDER test certificate.
Specifications | |
Exterior dimensions | |
Width (mm/inch) | 834 / 32.8 |
Height (inclusive feet/castors) (mm/inch) | 702 / 27.6 |
Depth (mm/inch) | 645 / 25.4 |
plus door handle, I-panel and exhaust duct (mm/inch) | 105 / 4.1 |
Wall clearance rear (mm/inch) | 100 / 3.9 |
Wall clearance side (mm/inch) | 160 / 6.3 |
Exhaust duct outer-Ø (mm/inch) | 52 / 2.1 |
Steam space volume (l/cu.ft.) | 158 / 5.6 |
Number of doors | 1 |
Interior dimensions | |
Width (mm/inch) | 600 / 23.6 |
Height (mm/inch) | 480 / 18.9 |
Depth (mm/inch) | 400 / 15.8 |
Interior volume (l/cu.ft.) | 115 / 4.1 |
Racks, chrome-plated (number standard/max.) | 2 / 6 |
Load per rack (kg/lbs.) | 20 / 44 |
Permitted total load (kg/lbs.) | 50 / 110 |
Weight of the unit (empty) (kg/lbs.) | 62 / 137 |
Temperature data | |
Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) | 300 / 572 |
Temperature variation 1) | |
at 70 °C (± °C) | 0.7 |
at 150 °C (± °C) | 1.8 |
at 300 °C (± °C) | 3.9 |
Temperature fluctuation (± °C) | 0.3 |
Heating up time 2) | |
to 70 °C (Min.) | 7 |
to 150 °C (Min.) | 30 |
to 250 °C (Min.) | 49 |
Recov. time after door was opened for 30 sec. 2) | |
at 70 °C (Min.) | 2 |
at 150 °C (Min.) | 8 |
at 300 °C (Min.) | 15 |
Electrical data | |
Housing protection acc. to EN 60529 | IP 20 |
Nominal voltage (±10 %) 50 / 60 Hz (V) | 230 |
Nominal power (kW) | 1.6 |
Energy consumption | |
at 70 °C (W) | 230 |
at 150 °C (W) | 544 |
at 300 °C (W) | 1100 |