- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results.
- Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F).
- Digital temperature setting with an accuracy of one degree.
- DS controller with integrated timer 0 to 99 h.
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm.
- Adjustable front ventilation flap slide and rear exhaust ø 50 mm (1.97 inch).
- With or without RS 422 Interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software.
- Units up to 115 liters are stackable.
- 2 chrome-plated racks included.
- BINDER test certificate.
Specifications | |
Exterior dimensions | |
Width (mm/inch) | 433 / 17.1 |
Height (inclusive feet/castors) (mm/inch) | 492 / 19.4 |
Depth (mm/inch) | 516 / 20.3 |
plus door handle, I-panel and exhaust duct (mm/inch) | 85 / 3.4 |
Wall clearance rear (mm/inch) | 100 / 3.9 |
Wall clearance side (mm/inch) | 100 / 3.9 |
Exhaust duct outer-Ø (mm/inch) | 52 / 2.1 |
Steam space volume (l/cu.ft.) | 36 / 1.3 |
Number of doors | 1 |
Interior dimensions | |
Width (mm/inch) | 222 / 8.7 |
Height (mm/inch) | 330 / 13.0 |
Depth (mm/inch) | 277 / 10.9 |
Interior volume (l/cu.ft.) | 20 / 0.7 |
Racks, chrome-plated (number standard/max.) | 2 / 3 |
Load per rack (kg/lbs.) | 12 / 26 |
Permitted total load (kg/lbs.) | 25 / 55 |
Weight of the unit (empty) (kg/lbs.) | 22 / 49 |
Temperature data | |
Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) | 300 / 572 |
Temperature variation 1) | |
at 70 °C (± °C) | 1.5 |
at 150 °C (± °C) | 2.5 |
at 300 °C (± °C) | 3.8 |
Temperature fluctuation at 70 °C (± °C) | 0.3 |
Heating up time 2) | |
to 70 °C (Min.) | 13 |
to 150 °C (Min.) | 24 |
to 250 °C (Min.) | 35 |
Recov. time after door was opened for 30 sec. 2) | |
at 70 °C (Min.) | 2.5 |
at 150 °C (Min.) | 5 |
at 300 °C (Min.) | 8 |
Air change data | |
Air change | |
at 70 °C (x/h) | 10 |
at 150 °C (x/h) | 13 |
at 300 °C (x/h) | 17 |
Electrical data | |
Housing protection acc. to EN 60529 | IP 20 |
Nominal voltage (±10 %) 50 / 60 Hz (V) | 230 |
Nominal power (kW) | 0.8 |
Energy consumption | |
at 70 °C (158 °F) (W) | 43 |
at 150 °C (302 °F) (W) | 148 |
at 300 °C (572 °F) (W) | 450 |
Accesories
ED 23 | |
Access port with silicone plugs, 10 mm (0.39 inch), 30 mm (1.18 inch), 50 mm (1.97inch) | ° |
Over temperature alarm, acoustic, can be switched off. Temperature limit can be set at the independent, adjustable temperature safety device class 2 | ° |
Independent adjustable temperature safety device, Class 3.1 (DIN 12880) | ° |
Analog temperature output, 4 – 20 mA, with 6 – pin DIN socket (output not adjustable) | ° |
Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C (302 °F) or at specified temperature with measuring protocol and certificate | ° |
Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature | ° |
Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature | ° |
Rack, chrome – plated or stainless steel | ° |
Shelf, perforated, stainless steel | ° |
Lockable door | ° |
FKM door gasket | ° |