Control range
- Temperature range from -70 to 150°C Humidity range from 25 to 95% RH
Performance
- Microprocessor PID control / Temperature calibration / Automatic tuning.
- Advanced performance of low and high temperature and humidity control, achieved by adapting the humidification tank exterior to the chamber, added benefit for long term test.
- Water recycling and filtering system reduce water consump-tion for longer operation
- Air-tight door closure with perfect sealed packing.
- Heat-resistant silicone packing completely blocks high heat leakage from the inner chamber.
- Two-point door latch also enhances tightness to door closure.
- Energy efficiency is achieved by optimization of the insulation system.
- Even if the unit is operating at max. temperature, the surface of the unit doesn’t allow high temperatures of over 60℃.
Safety
- Built in door lock with a key.
- Automatic shut off after overheat alarm.
- Low and empty water level alarm.
- Power supply leakage breaker.
- Over current protection.
- Open door alarm.
Convenience
- Three different volume sizes are available to suit any users’needs.
- Various volume sizes, available to suit any users’ need.
- Corrosion resistance stainless steel chamber and electric polished humidification tank.
- Maintenance of water supply are easily performed with a front water tank.
- Portable water container makes supplying water convenient.
- Water supply during operation is also available for long term test purposes.
- Direct water system provides user great convenience when performing long-term tests. (optional)
- Clear observation of samples.
- Tempered glass door with built in heating wire.
- Energy efficient, long life span LED lighting.
- Wider view, inner glass door is available for monitoring of samples. (optional)
- Cable port for introduction of external probes or wires.
- 50mm diameter cable port comes standard on the left side of the unit.
- Additional 50mm or 80mm diameter cable ports, custom ordered. (optional)
- Casters for easy mobility during installation or relocation.
- User friendly LCD controller.
-
- 5.7 inch colorful interactive touch type LCD controller.
- Set value and present value are clearly visible together on display.
- Operating progressvisible in graph mode allow for direct data interpretation.
- Built-in SD card drive.
- Trilingual support. (English / Chinese / Korean)
- Temperature and other operating parameters recorded with use of selectable recorder. (optional)
– Sufficient programmable control support.
-
Description | No | |
Programmable pattern capacity | 120 | |
Pattern repeat time | 999 | |
Max. segments per a pattern | 100 | |
Available max. segments* | 1200 | |
Programmable process time per a segment | 0 to 999hr 59 min 59 sec |
* Even though the max. segments per a pattern are 100, available total segments are up to 1200 not 12000, so you need to arrange the number of segment an patterns properly.
- Computer interface.
- RS-485 communication port enables connection of up to maximum 9 units. Simultaneous monitoring and operating of multiple units is an available feature.(optional)
- Program patterns are simple to set with the use of our software.
- Storage of program information, backup of value settings, recovery, temperature, humidity values.
- RS-232 communication port comes standard.
Models | TH-KH-180 | TH-KH-300 | TH-KH-408 | |||||
Chamber volume (L / cu ft) | 189 / 7.4 | 302 / 11.9 | 408 / 14.4 | |||||
Controller | Programmable, Touch screen type controller (5.7” color monitor) | |||||||
Temperature | Range (℃ / ℉) | -70 to 150 / -94 to 302 without humidity, 15 to 90 / 59 to 194 with humidity * Best programmable control range : -60 to 140 / -76 to 284 |
||||||
Fluctuation 1) (±℃ / ℉) at 40℃ / 75%RH | 0.3 / 0.54 | |||||||
Variation 1) (±℃ / ℉) at 40℃ / 75%RH | 0.5 / 0.90 | |||||||
Time | Heating – without humidity (min.) | 70 from -70℃ to 120℃ (-96℉ to 248℉) | ||||||
Cooling – without humidity (min.) | 80 from 20℃ to -70℃ (68℉ to -96℉) | |||||||
Humidity | Range (%RH) | 25 to 95 ※ See Feature page for more detailed humidity range check. |
||||||
Fluctuation 1) (±%RH) at 75%RH / 40℃ | 0.4 | |||||||
Variation1)(±%RH) at 75%RH / 40°C | 3.0 | |||||||
Refrigeration | System | Mechanical dual stage refrigerator system (air-cooled condenser) | ||||||
Refrigerant | R-507, R-23 | |||||||
Water supply | Capacity | 15L (0.53cu ft) – Cartridge type | ||||||
Water quality | pH6.2 to 7.2 / Electrical conductivity 20㎲ /cm to below | |||||||
No. of wire shelves (standard / max.) | 2 / 9 | 2 / 10 | 2 / 10 | |||||
Distance of between shelves (mm / inch) | 60 / 2.4 | 60 / 2.4 | 60 / 2.4 | |||||
Max. load per shelf (kg / lbs) | 20 / 44.1 | 25 / 55.1 | 25 / 55.1 | |||||
Permitted total load (kg / lbs) | 50 / 110.2 | 65 / 143.3 | 65 / 143.3 | |||||
Dimension (W×D×H) | Interior (mm / inch) | 500×630×600 / 19.7×24.8×23.6 | 600×630×800 / 23.6×24.8×31.5 | 600×850×800 / 23.6×33.5×31.5 | ||||
Exterior (mm / inch) | 1050x1615x1560 / 41.3×63.6×61.4 | 1150x1615x1760 / 45.3×63.6×69.3 | 1150x1835x1760 / 45.3×72.2×69.3 | |||||
Viewing windows (mm / inch) | 200×330 / 7.9×13 | |||||||
Net weight (kg / lbs) | 320 / 705.52 | 385 / 848.8 | 410 / 903.9 | |||||
Cable port | Basically one hole provided / Ø 50mm (2”), Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”) |
|||||||
Interface port | Basically RS-232C, [RS-485 : optional] | |||||||
Electrical requirements (380V, 50Hz) | 3ph / 10.1A | 3ph / 10.2A | 3ph / 10.2A | |||||
Cat.No. | AAHC3014K | AAHC3114K | AAHC3214K |