The T9/T10DCS Spectrophotometer incorporates dual monochromator technology making it well suited to even the most demanding of applications in all areas of UV-Visible Spectroscopy.
The optical design of both the T9DCS and T10DCS offer extremely low stray light characteristics (≤0.00004%T NaI,220 nm) which allows for an extensive photometric range (-8.0 – 8.0Abs). Measurements at deep ultra-violet wavelengths can also be achieved with use of Nitrogen purged optics.
The instrument can be optically configured to suit the needs of the sample by means of a continually adjustable slit for precise control of spectral resolution and beam size adjustment by means of an attenuating wheel.
Key Features
- True Double Beam Double monochromator Optics, and fully sealed to provide ultra-low stray light (0.00004%%T @220nm) and high photometric range (-8.0~8.0 Abs)
- 185– 900nm Wavelength range with Nitrogen purge
- Precise Wavelength accuracy is ensured by the integrated Mercury lamp for automatic correction
- Continuously selectable spectra bandwidth from 0.1-5nm
- Beam size continuously adjustable for variety of applications including ultra-volume cuvettes
- Large sample compartment suitable for all kinds of accessories, especially for 150mm integrating sphere
- Support WIFI communication in field operation
- The integrated Mercury Emissions Lamp with automatic correction function ensures precise wavelength accuracy
Model | T9DCS | T10DCS | |
Optical system | Double beam | Double beam | |
Scan speed | 1000 nm/min 1500 nm/min location speed |
1000 nm/min 1500 nm/min location speed |
|
Wavelength range | 185-900 nm | 185-900 nm | |
Wavelength accuracy | ±0.2 nm | ±0.2 nm | |
Spectral bandwidth | ±20 % (0.1 nm-5 nm continuously adjustable) | ±20 % (0.1 nm-5 nm continuously adjustable) | |
Photometric mode | Transmittance,Absorbance,Enegy and Concentration | Transmittance,Absorbance, Enegy and Concentration | |
Photometric range | -6.0 Abs-6.0 Abs | -8.0 Abs-8.0 Abs | |
Photometric accuracy | ±0.004 A@2.0 A ±0.003 A@1.0 A ±0.002 A@0.5 A ±0.3 % |
±0.004 A@2.0 A ±0.003 A@1.0 A ±0.002 A@0.5 A ±0.3 % |
|
Photometric reproducibility | ≤0.002 A@2.0 A ≤0.0008 A/1.0 A ≤0.0004 A/0.5 A ≤0.1 % |
≤0.002 A@2.0 A ≤0.0008 A/1.0 A ≤0.0004 A/0.5 A ≤0.1 % |
|
Photometric noise | 0 % Noise: ≤0.01 % 100 % Noise: ≤0.1 % |
0 % Noise: ≤0.01 % 100 % Noise: ≤0.1 % |
|
Baseline flatness | ±0.0008 Abs | ±0.0005 Abs | |
Baseline stability | 0.1 %/hr | 0.1 %/hr | |
Stray light | ≤0.0001 %T (NaI, 220 nm) ≤0.0001 %T (NaNO2, 360 nm) |
≤0.00004 %T (NaI, 220 nm) ≤0.00002 %T (NaNO2, 360 nm) |
|
Cell holder | Automatic 8 Cell Changer | Automatic 8 Cell Changer | |
Detector | Photo multiplier tube | Photo multiplier tube | |
Light source | D2 lamp,Tungstern lamp,Mercury lamp for WL correction | D2 lamp,Tungstern lamp,Mercury lamp for WL correction | |
PC interface | RS232, USB, Wifi | RS232, USB, Wifi | |
Software support | UV Win | UV Win | |
Power supply | Switchable 120-230 VAC 50-60 Hz | Switchable 120-230 VAC 50-60 Hz | |
Weight | 136 lb | 136 lb | |
Dimensions (Width, Depth, Height) |
36 x 25 x 12 in | 36 x 25 x 12 in |
MODEL | PART NO | DESCRIPTION |
T9DCS | T110+ | Double monochromater spectrometer with fixed cell holder, wavelength range 185-900nm with a user defined bandpass from 0.1-5.0nm in increments of 0.1nm. Stray light lower than 0.0001%T. PC controlled by USB or RS232 cable or WIFI |
T10DCS | T112+ | Double monochromater spectrometer with fixed cell holder, wavelength range 185-900nm with a user defined bandpass from 0.1-5.0nm in increments of 0.1nm. Stray light lower than 0.00004%T. PC controlled by USB or RS232 cable or WIFI |