The main advantage of the BINDER KBF LQC series is the patented light photometry for photostability tests. Two 3D spherical sensors control the light dosage of UV-A and visible light. Once the desired light dosage is achieved, the chamber switches off automatically.
Important characteristics
- Temperature range: 0 °C to 70 °C
- Temperature range with light: 10 °C to 60 °C
- Humidity range: 10 % to 80 % RH
- 3 positionable illumination cassettes with ICH-compliant UV/Vis light source
- Independent light-dose control of UV-A and visible light with spherical sensors
- APT.line™ preheating chamber technology
- Humidity regulation with capacitive humidity sensor and vapor humidification
- Inner chamber made of stainless steel
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Tight-sealing inner door made of safety glass (ESG) with special TIMELESS coating
- 3 stainless steel racks
- Access port with silicone plug, 30 mm, left
- 4 stable castors, two with brakes
- Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
- Computer interface: Ethernet
- Door heating
Designation | KBFLQC720-230V¹ | KBFLQC720UL-240V¹ |
Option model | Standard | Standard |
Order number | 9020-0334 | 9020-0335 |
Measures | ||
Interior volume [L] | 700 | 700 |
Net weight of the unit (empty) [kg] | 374 | 374 |
permitted load [kg] | 150 | 150 |
Load per rack [kg] | 45 | 45 |
Doors | ||
Inner doors | 2 | 2 |
Unit doors | 2 | 2 |
Internal Dimensions | ||
Width [mm] | 973 | 973 |
Height [mm] | 1250 | 1250 |
Depth [mm] | 576 | 576 |
Outer dimensions | ||
Width net [mm] | 1250 | 1250 |
Height net [mm] | 1925 | 1925 |
Depth net [mm] | 890 | 890 |
Wall clearance sidewise [mm] | 200 | 200 |
Wall clearance back [mm] | 100 | 100 |
Environment-specific data | ||
Energy consumption at 40 °C and 75 % RH [Wh/h] | 2350 | 2350 |
Sound-pressure level [dB(A)] | 53 | 53 |
Fixtures | ||
Number of shelves (std./max.) | 3/12 | 3/12 |
Number of illumination cassettes (std./max.) | 3/3 | 3/3 |
Performance Data Temperature | ||
Temperature range without illumination cassettes [°C] | 0…70 | 0…70 |
Temperature range with 100% illumination [°C] | 10…60 | 10…60 |
Max. heat compensation at 40 °C with illumination [W] | 1000 | 1000 |
Performance Data Climate | ||
Temperature range without illumination cassettes [°C] | 10…70 | 10…70 |
Temperature range with 100% illumination [°C] | 20…60 | 20…60 |
Temperature variation with illumination at 25 °C and 60 % RH [± K] | 1.2 | 1.2 |
Temperature variation with illumination at 40 °C and 75 % RH [± K] | 1.2 | 1.2 |
Temperature fluctuation with illumination at 40 °C and 75 % RH [± K] | 0.2 | 0.2 |
Temperature fluctuation with illumination at 25 °C and 60 % RH [± K] | 0.2 | 0.2 |
Humidity range without illumination [% RH] | 10…80 | 10…80 |
Humidity range with illumination cassettes, with illumination [% RH] | 10…75 | 10…75 |
Humidity fluctuation with illumination at 40 °C and 75 % RH [± % RH] | 2 | 2 |
Humidity fluctuation with illumination at 25 °C and 60 % RH [± % RH] | 2 | 2 |
Humidty recovery time after 30 s door open at 40 °C and 75 % RH with illumination [min] | 5 | 5 |
Humidty recovery time after 30 s door open at 25 °C and 60 % RH with illumination [min] | 1 | 1 |
Lightdata per Illumination Cassettes | ||
ICH compliant illumination for photo stability testing [lx]* | 9000 | 9000 |
ICH compliant illumination for photo stability testing [W/m²]* | 1.5 | 1.5 |
Electrical data | ||
Rated Voltage [V] | 200…230 | 200…240 |
Power frequency [Hz] | 50 | 50/60 |
Nominal power [kW] | 0 | 3.5 |
Unit fuse [A] | 16 | 16 |
Phase (Nominal voltage) [ph] | 1 | 1 |