Climate chambers with illumination and humidity The BINDER KBWF series shines with its homogeneous light distribution with natural illumination from above. It creates consistent climate conditions – even with illumination
Important characteristics
- Temperature range: 0 °C to 70 °C
- Temperature range with light: 10 °C to 60 °C
- Humidity range: 10 % to 80 % RH
- 2 positionable illumination cassettes, each with 5 fluorescent tubes
- APT.line™ preheating chamber technology
- Adjustable fan speed
- Humidity regulation with capacitative humidity sensor and vapor humidification
- Intuitive touchscreen controller with time-segment and real-time programming with visual alarm
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Tight-sealing inner door made of safety glass (ESG)
- Avoidance of glass corrosion by special TIMELESS coating
- Display via color LCD monitor
- 2 stainless steel racks
- Access port with silicone plug, 30 mm, left
- Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
- Computer interface: Ethernet
- Door heating
Designation | KBWF240-230V¹ | |
Option model | Standard | |
Order number | 9020-0336 | |
Measures | ||
Interior volume [L] | 247 | |
Net weight of the unit (empty) [kg] | 214 | |
permitted load [kg] | 100 | |
Load per rack [kg] | 30 | |
Doors | ||
Inner doors | 1 | |
Unit doors | 1 | |
Internal Dimensions | ||
Width [mm] | 650 | |
Height [mm] | 785 | |
Depth [mm] | 485 | |
Outer dimensions | ||
Width net [mm] | 925 | |
Height net [mm] | 1465 | |
Depth net [mm] | 800 | |
Wall clearance sidewise [mm] | 100 | |
Wall clearance back [mm] | 100 | |
Environment-specific data | ||
Energy consumption at 37 °C and 75 % RH [Wh/h] | 760 | |
Sound-pressure level [dB(A)] | 53 | |
Fixtures | ||
Number of shelves (std./max.) | 2/7 | |
Number of illumination cassettes (std./max.) | 2/2 | |
Performance Data Temperature | ||
Temperature range without illumination cassettes [°C] | 0…70 | |
Temperature range with 100% illumination [°C] | 10…60 | |
Performance Data Climate | ||
Temperature range without illumination cassettes [°C] | 10…70 | |
Temperature range with 100% illumination [°C] | 10…60 | |
Humidity range without illumination [% RH] | 10…80 | |
Humidity range with illumination cassettes, with illumination [% RH] | 10…75 | |
Lightdata per Illumination Cassettes | ||
Daylight tubes [W/m²]* | 24 | |
Daylight tubes [lx]* | 9000 | |
Fluora® growth lamps [W/m²]* | 23 | |
Fluora® growth lamps [lx]* | 7500 | |
Arabidopsis lamps [W/m²]* | 32 | |
Arabidopsis lamps [lx]* | 11000 | |
Electrical data | ||
Rated Voltage [V] | 200…240 | |
Power frequency [Hz] | 50 | |
Nominal power [kW] | 0 | |
Unit fuse [A] | 16 | |
Phase (Nominal voltage) [ph] | 1 |