- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results..
- Temperature range 5 °C (32 ºF) above ambient temperature up to 300 °C (572 °F).
- Digital temperature setting with an accuracy of one degree.
- DS controller with integrated timer 0 to 99 h.
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm.
- Adjustable front ventilation flap slide and rear exhaust ø 50 mm (1.97 inch).
- With or without RS 422 Interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software.
- Units up to 115 liters are stackable.
- 2 chrome-plated racks included.
- BINDER test certificate.
Specifications | |
Exterior dimensions | |
Width (mm/inch) | 1234 / 48.6 |
Height (inclusive feet/castors) (mm/inch) | 1022 / 40.2 |
Depth (mm/inch) | 765 / 30.1 |
plus door handle, I-panel and exhaust duct (mm/inch) | 85 / 3.4 |
Wall clearance rear (mm/inch) | 100 / 3.9 |
Wall clearance side (mm/inch) | 160 / 6.3 |
Exhaust duct outer-Ø (mm/inch) | 52 / 2.1 |
Steam space volume (l/cu.ft.) | 457 / 16.2 |
Number of doors | 2 |
Interior dimensions | |
Width (mm/inch) | 1000 / 39.4 |
Height (mm/inch) | 800 / 31.5 |
Depth (mm/inch) | 500 / 19.7 |
Interior volume (l/cu.ft.) | 400 / 14.3 |
Racks, chrome-plated (number standard/max.) | 2 / 10 |
Load per rack (kg/lbs.) | 35 / 77 |
Permitted total load (kg/lbs.) | 90 / 199 |
Weight of the unit (empty) (kg/lbs.) | 125/276 |
Temperature data | |
Temperature range, 5 °C (41 °F) above ambient up to (°C / °F) | 300 / 572 |
Temperature variation 1) | |
at 70 °C (± °C) | 1.7 |
at 150 °C (± °C) | 3 |
at 300 °C (± °C) | 5.0 |
Temperature fluctuation at 70 °C (± °C) | 0.3 |
Heating up time 2) | |
to 70 °C (Min.) | 49 |
to 150 °C (Min.) | 62 |
to 250 °C (Min.) | 74 |
Recov. time after door was opened for 30 sec. 2) | |
at 70 °C (Min.) | 4 |
at 150 °C (Min.) | 20 |
at 300 °C (Min.) | 24 |
Air change data | |
Air change | |
at 70 °C (x/h) | 11 |
at 150 °C (x/h) | 10 |
at 300 °C (x/h) | 9 |
Electrical data | |
Housing protection acc. to EN 60529 | IP 20 |
Nominal voltage (±10 %) 50 / 60 Hz (V) | 400 / 208 (3N) |
Nominal power (W) | 3400 |
Energy consumption | |
at 70 °C (W) | 201 |
at 150 °C (W) | 672 |
at 300 °C (W) | 1000 |
Accesories
ED 400 | |
Access port with silicone plugs, 10 mm (0.39 inch), 30 mm (1.18 inch), 50 mm (1.97 inch), 100 mm (3.94 inch) | ° |
Anti – slip rubber pads for safe stacking (4 pieces) | ° |
Over temperature alarm, acoustic, can be switched off. Temperature limit can be set at the independent, adjustable temperature safety device class 2 | ° |
Independent adjustable temperature safety device, Class 3.1 (DIN 12880) | ° |
Analog temperature output, 4 – 20 mA, with 6 – pin DIN socket (output not adjustable) | ° |
Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C (302 °F) or at specified temperature with measuring protocol and certificate | ° |
Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature | ° |
Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature | ° |
Rack, chrome – plated or tainless steel | ° |
Shelf, perforated, stainless steel | ° |
Lockable door | ° |
FKM door gasket | ° |
2 doors with 1 window each (350 x 240 mm / 13.78 x 9.45 inch) and interior lighting, 30 W . | ° |