Cabinets are as attractive as they are durable. Exteriors are painted light gray and have a hard, scratch-resistant hammer finish. Doors open with high-impact thermoplastic handles. Cabinets have heavy steel double-wall construction. Work space is insulated from the outer cabinet with one inch of high-density mineral wool, and interiors are made of corrosion-resistant aluminized steel.
Controls
The digital control combines the features of the analog model but offers the ease of temperature setting and the stability of a full PID microprocessor that accurately maintains settings within +/- 0.5o C (1o F), even in varying ambient or power supply conditions. This feature helps to confirm stability or indicate any control malfunction or power loss throughout a process period. The digital controller features large LED’s that continuously display set temperature and process temperature, as well as a setting lock mode that provides protection against accidental or inadvertent adjustment
Heating Elements
Energy-efficient, low-watt density, incoloy sheathed elements are engineered into a compact design for quick run-up and recovery times. Temperature uniformity is greatly improved by a perforated heat shield which absorbs radiant heat and distributes it more evenly throughout the product chamber.
GCE and AFE Models
Quincy Lab uses only quality UL and CSA recognized components in all ovens. The ‘GCE’ series gravity convection ovens are designed to meet the laboratory needs of industry, research organizations, and schools. Well-crafted and versatile, they are used for part drying, baking, curing, sterilizing, evaporating, heat treating, annealing, and testing. The ‘AFE’ models offer forced-air circulation allowing for uniform distribution of heat throughout the chamber, which also facilitates dehydration and evaporating. The GCE and AFE series ovens have temperature ranges of: ambient +25o F minimum, to 450o F/232o C maximum, and are competitively priced and offer exceptional value and reliability.
SPECIFICATIONS | MODEL 10GCE | MODEL 20GCE | MODEL 30GCE | MODEL 40GCE | |
INTERIOR DIMENSIONS | |||||
INCHES W x H x D | 12x10x10 | 13x13x13 | 18×15.6×12 | 18×21.8×14 | |
(CM) W x H x D | 30.5×25.4×25.4 | 33x33x33 | 45.7×39.6×30.5 | 45.7×55.4×35.6 | |
EXTERIOR DIMENSIONS | |||||
INCHES W x H x D | 14×17.5X12.3 | 15×21.5×15.3 | 20×25.5×14.3 | 20×31.5X16.3 | |
(CM) W x H x D | 35.6×44.5×31.2 | 38×54.6×38.9 | 50.8×64.8×36.2 | 2 50.8x80x41.4 | |
CAPACITY | |||||
CUBIC FEET | 0.7 | 1.27 | 2.0 | 3.0 | |
(LITERS) | 19.8 | 36 | 56.6 | 85 | |
TEMPERATURE RANGE | |||||
AMBIENT +25F TO F / C | 450/232 | 450/232 | 450/232 | 450/232 | |
SHELVES (1″ CENTERS) | |||||
MAXIMUM PER UNIT | 10 | 13 | 16 | 22 | |
MAXIMUM POUNDS / SHELF | 35 | 35 | 35 | 35 | |
ELECTRICAL* | |||||
VOLTS/AMPS | 115/5.2 | 115 / 6.3 | 115/10.5 | 115/12.5 | |
WATTS | 600 | 750 | 1200 | 1500 | |
PLUG/NEMA | 5-15P | 5-15P | 5-15P | 5-15P | |
WEIGHT | |||||
SHIPPING | 42.6 | 57 | 76.2 | 91.5 | |
STAND ALONE | 31.4 | 43.5 | 57.2 | 72.5 |
SPECIFICATIONS | MODEL 10AF | MODEL 20AF | MODEL 30AF | MODEL 40AF | |
INTERIOR DIMENSIONS | |||||
INCHES W x H x D | 12×8.25×10 | 13x11x13 | 18×14.2×12 | 18×19.8×14 | |
(CM) W x H x D | 30.5x21x25.4 | 33x28x33 | 45.7x36x30.5 | 45.7×50.3×35.6 | |
EXTERIOR DIMENSIONS | |||||
INCHES W x H x D | 14×20.5X12.3 | 15x25x15.3 | 20x29x14.3 | 20x35X16.3 | |
(CM) W x H x D | 35.6x52x31.2 | 38×63.5×38.9 | 50.8×73.7×36.2 | 50.8x89x41.4 | |
CAPACITY | |||||
CUBIC FEET | 0.6 | 1.14 | 1.83 | 2.86 | |
(LITERS) | 17 | 32.3 | 51.8 | 81 | |
TEMPERATURE RANGE | |||||
AMBIENT +25F TO F / C | 450/232 | 450/232 | 450/232 | 450/232 | |
SHELVES (1″ CENTERS) | |||||
MAXIMUM PER UNIT | 8 | 11 | 14 | 20 | |
MAXIMUM POUNDS / SHELF | 35 | 35 | 35 | 35 | |
ELECTRICAL* | |||||
VOLTS/AMPS | 115 / 6.6 | 115 / 8.8 | 115 / 12.5 | 115 / 12.5 | |
WATTS | 800 | 1050 | 1500 | 1500 | |
PLUG/NEMA | 5-15P | 5-15P | 5-15P | 5-15P | |
WEIGHT | |||||
SHIPPING | 50.4 | 65.7 | 84.2 | 98.5 | |
STAND ALONE | 37.2 | 58.7 | 64.2 | 80.8 |
**PERFORMANCE CHARACTERISTICS |
MODEL 10GCE |
MODEL 20GCE |
MODEL 30GCE |
MODEL 40GCE |
MODEL 10AFE |
MODEL 20AFE |
MODEL 30AFE |
MODEL 40AFE |
|
CONTROL STABILITY | |||||||||
Typically +/- C/F | 0.5° / 1.0° | 0.5° / 1.0° | 0.5° / 1.0° | 0.5° / 1.0° | 0.5° / 1.0° | 0.5° / 1.0° | 0.5° / 1.0° | 0.5° / 1.0° | |
TIME TO TEMPERATURE | |||||||||
AMBIENT TO MAX | 43 Min | 52 Min. | 37 Min. | 42 Min. | 43 Min. | 44 Min. | 38 Min. | 48 Min. | |
RECOVERY @300 °F | |||||||||
DOOR OPEN 15 SEC | 9 Min | 11 Min. | 10 Min. | 10 Min. | 10 Min. | 10 Min. | 11 Min. | 13 Min. | |