Contamination ControlExtensive use of copper in the CO2 sample port, humidity reservoir, and heated CO2 feed line adds reassurance that foreign microbes will not affect test results. Cleanup is a breeze with the all stainless steel chamber, coved corners, and autoclavable door gasket. Optional copper shelves are available for additional contamination control.
SCO6AD Unit Specification Chart | SCO6AD | |
Interior Volume, Nominal | 5.9 cu.ft. (167 liters) | |
Interior Dimensions W x D x H | 20.2″ x 20.0″ x 25.5″ 513 x 508 x 647 mm |
|
Exterior Dimensions(w x d x h) | 28.5″ x 30.8″ x 39.5″ 724 x 782 x 1003 mm |
|
Interior Construction | 300 series stainless steel | |
Exterior Construction | 20 gauge steel, powder coated | |
Shelves (See Accessories)* | 3 standard, 8 total | |
Maximum Weight Per Shelf* | 35 lb / 15.8 kg | |
Permitted Total Load | 105 lb / 47.6kg |
Order Model Number, Voltage Specific | SCO6AD | ||||||
SCO6AD | SCO6AD-2 | ||||||
Electrical, 50/60Hz, AC, 1Ø | |||||||
Voltage | 110V-120V | 220V-240V | |||||
Full Load Amps | 12 | 6.0 | |||||
Nominal Power (watts @37°C) | 130 | 110 | |||||
Recommended Breaker, Amps | 15 | 15 | |||||
Power Cord Supplied | NEMA5-15P | EU1-16P | |||||
CO2 Range | 1-20% | 1-20% | |||||
Recovery CO2 at 5%* | < 5 min. | < 5 min. | |||||
Sensor Type | Infrared | Infrared | |||||
Temperature Uniformity | |||||||
At 37°C | ± 0.25°C | ± 0.25°C | |||||
Humidity Range Passive (Ambient 95%) | ✓ | ✓ | |||||
Temperature Range Recovery 37°C* | |||||||
Recovery to 37°C after 30 sec. Door Opening | 6.0 min. | 6.0 min. |